THOR 5 - Thermal Test for Displays - Alrad
THOR 5 - Thermal Test for Displays - Alrad

THOR 5 - Thermal Test for Displays

THOR 5  Thermal Test for Displays

The design is modular and allows a use of the bench with or without the temperature chamber.  This complete measurement station is an accurate and fully automated solution which delivers complete test results reports with the automation software provided.

OVERVIEW

THOR5 system is an automated measurement station which offers easily an evaluation of displays up to 20” in specific temperature environment.

The design is modular and allows a use of the station with or without the temperature chamber. This complete measurement station is an accurate and automated solution which delivers complete test results report with the automation software provided.

FEATURES
  • Fully automated +/- 80 ̊ or +/- 88 ̊ viewing angle test station for displays
  • All included station: Measuring system EZContrast and its valets, thermic chamber and automated XYZ stage
  • High flexibility, accuracy and ease of use
  • Resolution: 0.08°/CCD pixel

APPLICATIONS

  • Viewing angle analysis for QC or R&D
  • Automated stage for fast and easy testing
  • Environmental impact on displays optical characteristics checked with the use of temperature chamber from -40 ̊ to + 85 ̊C
  • Reflective properties and/or spectral characteristics of displays available as options

Thermal Image

Specifications   THOR5
Viewing angle Incident angle
Azimuth angle
0-80° or 0-88°
0-360°
Measuring area Maximum diameter 2mm
Optimum distance Ensure light coming from same spot at any angle 3.7mm or 1mm
Accuracy Luminance
Chromaticity (x,y) RMS
±3%
0.002
Performances Angular resolution
Peltier cooled
Resolution
A/D converter
Angular optical resolution
0.08°/CCD pixel
Yes
6M
14-16 bits
±0.3°(*)
Chromaticity Color Filters 6
Measurement Modes Luminance
Chromaticity
Yes
Yes
Luminance Range Minimum (cd/m2)
Maximum (cd/m2) - on option with neutral density
0.001
1000000
Short-term repeatability Luminance
Chromaticity (x,y) RMS
±0.05%
0.0001
Measurement time Luminance
Chromaticity
< 10s
<20s
Thermic Chamber Test space volume (liter)
Test space dimensions (mm)
External dimensions (mm)
990
height: 500; width: 490; depth: 400
height: 1190; width: 780; depth: 930
Manual XYZ Stage Courses X
Courses Y
Courses Z
Removable Sampler Holder
Accuracy
±300mm
±350mm
±500mm
5kg max charge
0.1mm
Specifications can be changed without notice