EZLite-HxS
The EZLite-HxS series gives spectral data within a+/- 60° viewing cone, and is very fast. A full spectral map is made in less than 3mn. With 1s band pass filters regularly distributed in the middle range, the EZLite-HxS is the solution for angular spectral analysis.
Features:
- Non-contact measurement
- Fast spectral measurements
- Radiance, luminance and color data
- High accuracy and reliability
- Cooled CCD sensor
Applications:
- Physico-realistic simulation
- All kind of displays performances characterization in details
- Grey levels analysis
Specifications | EZLite-HxS | |
---|---|---|
Viewing Angle | Incident angle Azimuth angle |
±60° 0-360° |
Measuring area | Diameters | 2mm, 1mm, 500μm, & 300μm |
Density | Set of neutral density | |
Polarization | Optimal | 3 polarizers: 0, 45 & 90° and 2 wave-plates |
Optimum distance | (To ensure the light is coming from the same spot at any angle) | 4,5mm |
Spectral specs | Standard Optional Spectral data extraction |
15 bandpass filters on the visible range 400-700nm 2 additional badpass filters between700 and 900nm Interpolation with steps between 1 and 5nm |
Measurement time | Radiance with full resolution Radiance with half resolution Polarization with full resolution |
<6mn <3mn <15mn |
Accuracy | Wavelength resolution (nm) Wavelength accuracy (nm) Stray light (%) Angular resollution (deg) Radiance (W/Str/m2/nm) Chromaticity Ellipticity & polarization direction Polarization degree |
20 1 <0.1 0.75 ±5% 0.005 (for any stimulus) ±2° up to 60° ±2% up to 60° |
Repeatability | Radiance Luminance Chromaticity |
±0.05% ±0.05% 0.001 |
Reflective options | Only Diffused | |
Use conditions | Temperature range Humidity range |
0 to 30°C 0-85% non condensing |
Interface | USB 2.0 | |
Power | AC adapter (100-240V 50/60Hz) | |
Current consumption | 90W | |
Weight | 10Kg | |
Specifications can be changed without notice |